PRODUCTS LIST
Home > Product Show
|
infusion pump with piggy function (MP1000)
Product Numbers: MP1000
Product description: A. Piggyback function for infusion of extra bag of other but compatible drug during or before primary infusion B. Small, lightweight and compact with Rate mode and Drop mode C. Easy panel with number keys and function keys D.Higher precision by3-rate accuracy calibration E. 0~9 tube channels for storage of different brands of IV sets
|
|
INTRODUCTION
easy to carry and cost-effective for transport clamp /hook for flexible fixta
Accuracy error rate
|
±5%(±3%after proper calibration)
|
Applicable IV sets
|
All brands of IV sets of national standard (tubing diameter: 3.5-4.5mm)
|
Rate range
|
1~600 ml/h
|
VTBI
|
1-9999ml
|
Prime/Bolus rate
|
600ml/h
|
Piggy function
|
Piggy rate: 1-600ml/h, Piggy VTBI: 1-9999ml
|
KVO rate
|
Upon ‘over’ alarm, the pump continues at 6ml/h to keep vein open.
|
Power
|
AC100~240V, 50/Hz±1Hz
|
Fuse
|
T 2AL 250V
|
Built-in battery
|
Rechargeable lithium ion battery, 8.4V, 1600mAh
|
Duration of battery
|
Run continuously for over 2 hours at the fastest rate (600 ml/h) after being fully charged.
|
Battery charging
|
The pump starts charging the battery automatically after being connected to AC power. It needs 8~12 hours (power-on) to get battery fully charged.
|
Displayed information
|
Rate, VTBI, VI, air, occl, over, AC / battery symbol, L.batt., Er, door open, tube, occlusion level, piggy indicator light
|
Alarm functions
|
over, L.batt., air, occl, er, door open
|
Occlusion gate
|
“Occl” alarm shall be given when tubing pressure reaches 0.08Mpa~0.16 Mpa. The occlusion dose is ≤0.8ml. The responsive time is respectively: 2~3 seconds at higher rate (400ml/h)
5~7 seconds at lower rate (200ml/h)
35~50 seconds at very low rate (25ml/h)
|
Shell size
|
130x130x80 mm (LxHxD)
|
Net weight
|
≤1.15kg
|
Classification
|
Class II, type BF, waterproof level: IXP3
|
Shell material
|
Environmental-friendly ABS plastic
|
Operation conditions
|
Ambient temperature: 5~40℃, atmospheric pressure: 70~106kPa,relative humidity: 20%~80%
|
Storage conditions
|
Ambient temperature: -15~50℃, atmospheric pressure: 70~106kPa, relative humidity: 20%~80%
|
Electricity safety
|
GB 9706.1-2007,GB 9706.27-2005
|
|