● voice alarm & audio-visual alarm
● upstream & downstream pressure sensor
● occlusion release and auto re-run
● needle-off detection
● double CPU
● number key for easy setting
● multifunctional for flexible use
● fixed by either hook or clamp
● small size, light weight
● Piggy-back function
●Light and compact (1.15kg)
●open system—accept all standard IV sets
infusion accuracy |
± 5% (± 3% with infusion set of good quality after proper calibration) |
Applicable IV set |
15, 20, 60drops/ml, tubing diameter: 3.5-4.5mm |
Infusion rate |
0.1-1200ml/h
increment selectable: 0.1ml/h, 1ml/h, 10ml/h or 100ml/h |
VTBI |
0.1-9999ml, increment: 0.1ml/h |
Accumulated volume |
0.1-9999ml |
KVO rate |
5ml/h |
Bolus rate |
1-1200ml/h (adjustable); default: 600ml/h |
Purge rate |
adjustable,default: 600ml/hr
|
Air-in-line detection |
mini. size of air bubble: 50ul (Levels 1, 2, 3, 4 adjustable); default: Level 1 |
Occlusion pressure |
double pressure sensor, up and down occlusion alarm |
Waterproof level |
IPX3 |
AC power |
100-240V 50/60Hz |
Built-in battery |
Li_Polyme 7.4V 1600mAh. Charging time: 8-12h with power on. Running time:5-6 h at rate 25ml/h, environment temperature 25C, after being fully charged. |
Power consumption |
25VA |
On-board battery/DC |
DC 12V ± 1V |
Working conditions |
Environment temperature 5C~40C
Relative humidity: 10-90% (free of frost)
Atmosphere pressure 86kPa~106kPa |
Shell size |
130(w)x 130(h)x 80(d)mm |
Net weight |
≤ 1.1kg |