Product Numbers: 500II
Product description: syringe pump with rate mode, time mode and body-weight mode. drug name library. accept all brands of national standard syringes 5ml, 10ml, 20/30ml, 50/60ml. easy to operate and good in accuracy.
|
|
INTRODUCTION
Function and Features of Syringe Pump
1). 3 kinds of modes: Rate Mode, Time Mode, Body Weight Mode.
2). Double CPU: Making injection process safer and more reliable.
3). Accurate & sensitive pressure sensor: Detecting occlusion pressure accurately.
4). Injection time, dose, solution: Easy to be checked during injection process.
5). Precision step motor: No pulsating wave even during low-rate injection.
6). KVO & Bolus functions: Rate selectable according to actual clinical requirement.
7). Excellent injection operation: Accuracy ensured even at small-volume injection.
8). Unique injection supervision system: Ensures safe injection and reduces the nurse's workload.
9). Open system for syringes: Accept all brands of syringes of national standard from 5ml, 10ml, 20ml, 30ml to 50ml, which can be self-diagnosed and displayed on LCD display.
10). Alarms: Audio-visual alarms given in case of malfunction or operation error with machine stopping at the same time automatically.
11). Preset volume: Automatically saved in case of unpredictable switch-off, thus ensuring accuracy of injection volume.
12). Drug name library: As many as 200 drug names can be stored in the machine and can be checked on LCD display during operation.
Dimension |
288x130x122mm (outer shell: ABS plastic) |
|
Power supply |
50/60Hz AC100V~240V |
|
Weight |
<3.0kg |
Battery |
Rechargeable lithium polymer battery, 7.4V, 1600mA |
|
Selectable modes |
Rate mode, Time mode, Body Weight mode |
|
Fuse |
220V2Ax2, 12V2Ax2 |
|
Applicable syringes |
Accept all brands of standard syringes:
5ml, 10ml, 20ml, 30ml, 50ml |
Electrical safety |
Comply with the stipulation of GB9706.1-1995, EN60601-1 and UL60601-1 |
|
Accumulated injection volume |
0~9999.9ml |
Maximum power consumption |
25VA, can run for 5h after being recharged |
|
Injection accuracy |
±2% |
|
Feasible syringe
rate range |
5ml: 0.1~100ml/h 10ml: 0.1~200ml/h
20ml: 0.1~400ml/h 30ml: 0.1~600ml/h
50ml: 0.1~1500ml/ h |
Batteryrecharge |
When the pump is connected to the AC power supply, turn on the power and the battery will automatically recharge (about 8-14 hours to recharge fully). |
|
Displayed Information |
rate, volume limit, syringe size, pressure value, battery capacity, unit of medical dosage, bed No. AC power indicator, accumulated volume, drug name |
|
Alarm function |
OVER, NEAR, Occlusion, Low battery, Syringe disengaged, Syringe abnormal, Control abnormal, Internal malfunction |
|
Electromagnetic compatibility |
Comply with the stipulation of IEC60601-1-2 (1993 edition) EN60601-2-24 |
|
Bolus rate |
5ml: 100ml/h, 10ml: 200ml/h,
20ml: 400ml/h,
30ml: 400~600ml/h (adjustable)
50ml: 400~1500ml/h (adjustable) |
|
Operating conditions |
Environment temperature 5ºC~45ºC, air pressure 70~106kpa, relative humidity 30%~90% |
|
KVO rate |
0.1ml/h~5.0ml/h(selectable) |
Storage |
Environment temperature -15ºC~50ºC, air pressure 70~106kPa, relative humidity 30%~90% |
|
Occlusion value |
400mmHg~900mmHg |
Water proof |
IPX1 |
|
Time mode |
Vol limit: 0.1~999.9ml Time: 0~2000 min |
Classification |
Class II, type BF |
|
Drug name library |
200 drug names can be stored in the machine. |
Certification |
CE0197 |
|
Body Weight Mode |
Drug limit: 0.1~9999.9(mg/ug), Solution limit: 0.1~999.9ml, Weight: 0.1~300.0kg, Dose: 0.1~9999.9 Selectable unit: mg/kg/h, ug/kg/min |
Applied standards |
Accord with the stipulation of YZB/Yue0482-2004 |
|
|