PRODUCTS LIST
Home > Product Show
|
Vet infusion pump (SM-MP1100) pro version
Product Numbers: MP1100 pro version
Product description: A. Small, lightweight and compact infusion pump with rate mode and drop mode. B. simple and clear panel with number keys easy and quick for setting. C. 3-rate calibration to ensure better accuracy D.
0~9 tube channels to store different brands of IV sets
|
|
INTRODUCTION
Vet Infusion Pump (Use for Veterinary)
light & compact,easy to carry and transport clamp /hook flexible for fixation
Accuracy error rate
|
±5%(±3%after proper calibration)
|
Applicable IV sets
|
All brands of IV sets of national standard (tubing diameter: 3.5-4.5mm)
|
Rate range
|
0.1~1200 ml/h
|
VTBI
|
1-9999ml
|
Prime/Bolus rate
|
600ml/h (adjustable)
|
KVO rate
|
Upon ‘over’ alarm, the pump continues at 6ml/h to keep vein open.
|
Power
|
AC100~240V, 50/Hz±1Hz
|
Fuse
|
T 2AL 250V
|
Built-in battery
|
Rechargeable lithium ion battery, 8.4V, 1600mAh
|
Duration of battery
|
Run continuously for over 5 hours at the fastest rate (600 ml/h) after being fully charged.
|
Battery charging
|
The pump starts charging the battery automatically after being connected to AC power. It needs 8~12 hours (power-on) to get battery fully charged.
|
Displayed information
|
Rate, VTBI, VI, air, occl, over, AC / battery symbol, L.batt., Er, door open, tube, occlusion level
|
Alarm functions
|
over, L.batt., air, occl, er, door open
|
Occlusion gate
|
“Occl” alarm shall be given when tubing pressure reaches 0.08Mpa~0.16 Mpa. The occlusion dose is ≤0.8ml. The responsive time is respectively: 2~3 seconds at higher rate (400ml/h)
5~7 seconds at lower rate (200ml/h)
35~50 seconds at very low rate (25ml/h)
|
Shell size
|
130x130x80 mm (LxHxD)
|
Net weight
|
≤1.15kg
|
Classification
|
Class II, type BF, waterproof level: IXP3
|
Shell material
|
Environmental-friendly ABS plastic
|
Operation conditions
|
Ambient temperature: 5~40℃, atmospheric pressure: 70~106kPa,relative humidity: 20%~80%
|
Storage conditions
|
Ambient temperature: -15~50℃, atmospheric pressure: 70~106kPa, relative humidity: 20%~80%
|
Electricity safety
|
GB 9706.1-2007,GB 9706.27-2005
|
|