INTRODUCTION
Function and Features Innovative door-less design for easy operation Middle Detection and Ultrasound Detectors to double guarantee the correct IV set installation Anti-bolus system to reduce significantly bolus after occlusion sudden release Optional drip sensor and fluid heater Wireless data transmission with infusion supervision system
Technical Specifications Mechanism Peristaltic pump Operating Mode Flow Rate Mode Preset Volume 1-9999 ml Accumulated Infusion Volume 1-9999 ml Flow Rate Range 0.1 - 1200 ml/h Increment 0.1, 1,10, 100 (ml/h) Accuracy ± 5% KVO Rate 1-5 ml/h Bolus Rate 600 ml/h
Display Screen 2.7 inch B/W LCD Information Flow Rate and unit, Preset Volume and unit, Accumulated infusion volume and unit, Finish, Occlusion, Air Bubble, Bed No., AC Power indicator, Battery capacity, Battery charge indicator, IV set, etc. Status Indicator Stop, Run, Bolus, KVO
Alarm Type Air Bubble, Finish, Occlusion, No AC Power, Low Battery, Battery Empty, Reminder to operate, Communication Error (Err1), Operation Error (Err2), Drops Error (Err3), Handle Open (Err4) Air Bubble Level: 1-3 levels adjustable, turn on/off Occlusion Alarm Pressure Range: 3 levels: 60kPa, 100kPa, 140kPa (±20kPa), Level Low as default
Battery Type Lithium polymer battery Life Up to 4 hours continuous operation@25 ml/h flow rate
Environment Temperature 5 - 40 ℃ for operating, -20 - 50 ℃ for storage Air Pressure 86 - 106 kPa for operating, 50 - 106 kPa for storage Humidity 20 % - 80 % for operating, 10 % - 95 % for storage
Others Classification Class II, Type BF Dimension 120 mm(L) × 140 mm(D) × 195 mm(H) Power Supply 100 - 240 VAC, 50/60 Hz Waterproof level IPX1 |