PRODUCTS LIST
Home > Product Show
|
Infusion Pump (600II)
Product Numbers: 600II
Product description: infusion pump with rate mode, time mode and body-weight mode. rate range 1-2000ml/h. easy to operate and good in accuracy.
|
|
INTRODUCTION
Function and Features
1). Flow rate range: 0.1-2000 ml/h
2). 3 modes: Rate mode, Time mode, Weight mode
3). Unique door-free structure: Making it easier for user to handle and avoid problems of pump stuck and fluid leakage
4). Big LCD screen: Making it easier and more convenient to operate
5). Unique bedside infusion supervision function: Making patients and their relatives more assured while reducing nurses' workload.
6). Open system tor IV set: Accept all brands of standard IV sets with only a simple calibration to ensure accuracy. Can keep values for 3 brands/types of iv sets
7). Audio-visual alarms: Alarms given in case of air in line or occlusion with machine
Stopping work automatically.
8). KVO: Selectable, 1.0 ml/h-5.0 ml/h
9). Infrared sensor: Accurate air alarm given when detecting excessive air pressure in line.
10). Double CPU: Making infusion process safer and more reliable.
11). Alarm signals are given in case of malfunction or operation error and the machine stops operation automatically.
Dimension |
120×140×190 mm |
|
Accumulated infusion volume |
1.0~9999.0 ml |
Maximum weight |
< 3.0 kg |
Infusion accuracy |
±5% |
Displayed information |
Infusion rate, preset volume, accumulated infusion volume, battery capacity, AC power indicator, infusion completion, occlusion, air bubbles, bed No., pressure value, flow rate. |
Alarm function |
Over,occlusion,air bubbles,low battery,installation error of infusion pipe,machine malfunction,empty |
Fuse |
T 2AL 250V~ |
KVO rate |
1.0 ml/h-5.0 ml/h |
Maximum power consumption |
25VA, running more than 4 hours after being charged |
Selectable modes |
Rate mode, Time mode, Weight mode |
Infusion rate range |
1.0-2000 ml/h |
Rate mode |
1.0~2000.0 ml/h |
Bolus |
Bolus: 1000.0 ml/h when Rate is ≤1000.0 ml/h |
Time mode |
Volume limit: 0.1~999.9ml Time: 1~2000 minutes |
Operation conditions |
Environment temperature 5ºC ~ 40ºC, air pressure 86kPa~106kPa, relative humidity ≤80% |
Body Weight mode |
Drug limit: 1.0~999.9 mg, Volume limit: 0.1~999.9 ml. Weight: 0.1~300.0 kg, Dose: 0.1~9999.9
Selectable unit: mg/kg/h, ug/kg/min |
Electrical Safety |
Comply with EN60601-1 for medical equipment and EN60601-2-24 |
Battery |
Rechargeable lithium polymer battery, 7.4V, 1600 mA |
Applicable infusion sets |
All standard IV sets |
Storage |
Environment temperature-20ºC~55ºC, air pressure 50~106kPa, relative humidity ≤95% |
|